Abstract

An approach to radiosurgery treatment that can be readily adopted in most radiotherapy centers with linear accelerators is presented. In our institution, a Leksell-type of neurosurgical frame, a computed tomography scanner, locally fabricated cones, and 6 MV X-ray beams are used to perform radiosurgery treatments. Collimated arcs with dose distributions that conform to the shape of the lesion in the transverse and the sagittal planes are used. It is argued that the uncertainties in the localization of the isocenter within a lesion and the specifications of the size of the target volume do not justify high precision mechanical devices for most radiosurgery treatments.

Original languageEnglish
Pages (from-to)113-117
Number of pages5
JournalMedical Dosimetry
Volume18
Issue number3
DOIs
StatePublished - 1993