Comparison of skin dose calculated by the dose tracking system (dts) with a beam angular correction factor and that calculated by monte-carlo

  • SUNY Buffalo

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

Skin dose is dependent on the incident beam angle and corrections are needed for accurate estimation of the risk of deterministic effects of the skin. Angular-correction factors (ACF) were calculated and incorporated into our skin-dosetracking system (DTS) and the results compared to Monte-Carlo simulations for a neuro-interventional procedure. To obtain the ACF's, EGSnrc Monte-Carlo (MC) software was used to calculate the dose averaged over 0.5, 1, 2, 3, 4 and 5 mm depth into the entrance surface of a water phantom at the center of the field as a function of incident beam to skin angle from 90-10 degrees for beam field sizes from 5-15 cm and for beam energies from 60-120 kVp. These values were normalized to the incident primary dose to obtain the ACF. The angle of incidence at each mesh vertex in the beam on the surface of the DTS patient graphic was calculated as the complement of the angle between the normal vector and the vector of the intersecting ray from the tube focal spot; skin dose at that vertex was calculated using the corresponding ACF. The skin-dose values with angular correction were compared to those calculated using MC with a matching voxelized phantom. The results show the ACF decreases with decreasing incident angle and skin thickness, and increases with increasing field size and kVp. Good agreement was obtained between the skin dose calculated by the angular-corrected DTS and MC, while use of the ACF allows the real-time performance of the DTS to be maintained.

Original languageEnglish
Title of host publicationMedical Imaging 2021
Subtitle of host publicationPhysics of Medical Imaging
EditorsHilde Bosmans, Wei Zhao, Lifeng Yu
PublisherSPIE
ISBN (Electronic)9781510640191
DOIs
StatePublished - 2021
EventMedical Imaging 2021: Physics of Medical Imaging - Virtual, Online, United States
Duration: Feb 15 2021Feb 19 2021

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume11595

Conference

ConferenceMedical Imaging 2021: Physics of Medical Imaging
Country/TerritoryUnited States
CityVirtual, Online
Period02/15/2102/19/21

Keywords

  • Backscatter
  • Diagnostic x-ray dosimetry
  • Fluoroscopy
  • Interventional radiology
  • Monte carlo simulation
  • Patient dose
  • Radiation dose
  • Skin dose
  • X-ray

Fingerprint

Dive into the research topics of 'Comparison of skin dose calculated by the dose tracking system (dts) with a beam angular correction factor and that calculated by monte-carlo'. Together they form a unique fingerprint.

Cite this