Skip to main navigation Skip to search Skip to main content

Comparison of the imaging physics performance of a prototype flat panel detector with a 400 speed screen-film system

  • W. Huda
  • , K. Ogden
  • , M. L. Roskopf
  • , C. Rush
  • SUNY Upstate Medical University

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

The performance of a digital radiography system that included a prototype flat panel detector (StingRay) was compared with a 400 speed screen-film system. The flat panel detector consisted of a 500 μm thick CsI scintillator with an image matrix size of 3k2. The limiting spatial resolution of screen-film (∼4 line pairs/mm) was superior to that of the flat panel detector (∼2.5 line pairs/mm). The digital detector had an excellent linearity response (r2 = 0.997), a dynamic range of 20,000:1, and saturated at a radiation exposure of 60 mR. At exposures > 50 μR, the flat panel noise performance was dominated by quantum mottle. For radiation exposure level which produce film densities of ∼1.8, the low contrast performance of the flat panel detector was similar to that of the screen-film system. Changes in radiation exposure, however, significantly affected the performance of the flat panel detector, whereas the performance of screen-film was constant at film densities between 1.5 and 2.5. The digital imaging system produced images for review in ∼13 seconds, which is much faster than the minimum 90 second processing time of film. Raw image data sets acquired using the digital detector showed four individual sub-panels, an artifact that can be eliminated by processing. The flat panel detector investigated in this work offers a wide dynamic range, excellent linearity performance and the rapid availability of digital images. The digital detector will also permit a reduction in radiation exposure where the detection task is relatively easy, or when the radiation to the patient is of specific concern.

Original languageEnglish
Pages (from-to)198-208
Number of pages11
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume4320
DOIs
StatePublished - 2001

Keywords

  • Flat panel detector
  • Imaging physics
  • Screen-film
  • System performance

Fingerprint

Dive into the research topics of 'Comparison of the imaging physics performance of a prototype flat panel detector with a 400 speed screen-film system'. Together they form a unique fingerprint.

Cite this