Abstract
Questions regarding the feasibility of using electrical impedance tomography (EIT) to detect breast cancer may be answered by building a sufficiently precise multiple frequency EIT instrument. Current sources are desirable for this application, yet no current source designs have been reported that have the required precision at the multiple frequencies needed. We have designed an EIT current source using an enhanced Howland topology in parallel with a generalized impedance converter (GIC). This combination allows for nearly independent adjustment of output resistance and output capacitance, resulting in simulated output impedances in excess of 2 GΩ between 100 Hz and 1 MHz. In this paper, the theoretical operation of this current source is explained, and experimental results demonstrate the feasibility of creating a high precision, multiple frequency, capacitance compensated current source for EIT applications.
| Original language | English |
|---|---|
| Pages (from-to) | 509-516 |
| Number of pages | 8 |
| Journal | Physiological Measurement |
| Volume | 24 |
| Issue number | 2 |
| DOIs | |
| State | Published - May 2003 |
Keywords
- Current source
- Electrical impedance imaging
- Electrical impedance tomography
- Generalized impedance converter