Abstract
A gas chromatographic procedure was developed for the determination of methanol in small-volume whole blood samples. Samples (100-200 μl) were prepared by protein precipitation, with direct injection of the supernatant on a wide-bore capillary column. The recovery of methanol and acetonitrile (the internal standard) was approximately 90% and did not vary with sample volume. The assay was linear from 2 μg/ml (the limit of detection) through 1000 μg/ml and was highly reproducible (intra-day coefficient of variation <2.5%). Assay performance was assessed following exposure of rats to methanol. The results indicate that the present procedure is suitable for studies of methanol disposition in small rodent species.
| Original language | English |
|---|---|
| Pages (from-to) | 406-411 |
| Number of pages | 6 |
| Journal | Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences |
| Volume | 570 |
| Issue number | 2 |
| DOIs | |
| State | Published - Oct 4 1991 |