Digital tourniquets: A pressure study with clinical relevance

  • F. Peter Hixson
  • , Bruce B. Shafiroff
  • , Frederick W. Werner
  • , Andrew K. Palmer
  • SUNY Upstate Medical University

Research output: Contribution to journalArticlepeer-review

42 Scopus citations

Abstract

This study measured the pressures under the three most commonly used digital tourniquets—the Penrose drain, the rolled rubber glove, and the rubber band. A miniature pressure transducer and a digital strain indicator were used to measure pressures generated by the various tourniquets. The rolled rubber glove technique was highly consistent, irrespective of the clinical experience of the subject, and uniformly generated pressures of less than 500 mm Hg. Pressures generated by Penrose drains and rubber bands were highly variable and were significantly greater than 500 mm Hg. The so-called calibrated Penrose drain generated the highest pressures in the study. The relationship between tourniquet pressures and neurovascular injury in the human digit is not clearly defined.

Original languageEnglish
Pages (from-to)865-868
Number of pages4
JournalJournal of Hand Surgery
Volume11
Issue number6
DOIs
StatePublished - 1986