Abstract
Accurate measurement and control of flow properties are essential for microfluidic lab-chips. In this paper, we demonstrate an electrolytic-bubble-based approach to directly measure flow rate in microfluidic arteries. By simultaneously generating two gas bubbles electrochemically along a channel, we can measure the pressure difference (and thus flow rate) in real time. A prototype chip that measures flow rate was fabricated on silicon. The sensor was characterized for a functional microfluidic system with inlet pressure ranging from 108 to 135 kPa. The flow-meter performance was compared with computational fluid dynamics simulations and was also calibrated against other direct experiments. The impedance-based flow-rate measurements are easily achieved on a chip with a simple electronic circuit. The described approach can be integrated into any fluid circuit, particularly microfluidic channels that are too small to use off-shelf flow meters.
| Original language | English |
|---|---|
| Pages (from-to) | 1087-1094 |
| Number of pages | 8 |
| Journal | Journal of Microelectromechanical Systems |
| Volume | 16 |
| Issue number | 5 |
| DOIs | |
| State | Published - Oct 2007 |
Keywords
- Bubbles
- Impedance measurement
- Microelectromechanical devices
- Microsensors
Fingerprint
Dive into the research topics of 'Electrolytic-bubble-based flow sensor for microfluidic systems'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver