TY - GEN
T1 - Energy window optimization of PET detectors for SPECT imaging
AU - Yao, Rutao
AU - Cadorette, Jules
AU - Beaudoin, Jean François
AU - Lecomte, Roger
PY - 2013
Y1 - 2013
N2 - We developed add-on SPECT imaging capability on an animal PET scanner to implement a cost effective dual modality imaging system. Using PET detectors, designed for 511 keV photons, to acquire relatively low energy gamma rays from SPECT radionuclides has some inherent downsides, including the sub-optimal energy resolution and background radioactivity from 176Lu, a long-lived radioisotope of Lutetium that is present in most of the modern PET detector scintillators. To address these downsides, this study was carried out to optimize the detector energy window setting in SPECT imaging mode. A LabPET-8 scanner with a pinhole collimator insert was studied through a Monte Carlo simulation package. The energy spectra of detector crystal's intrinsic 176Lu, and a cylindrical imaging phantom's 99mTc emissions were simulated and assessed. The optimal energy window setting was derived by searching the lower and upper energy thresholds that maximize the noise-equivalent-count rate (NECR), an index adapted from its PET counterpart. The characteristics of 176Lu energy spectrum obtained from simulation matches that reported in the literature and reproduces the features of 176Lu energy spectra measured on the LabPET-8 scanner. The optimal energy window setting derived from the simulation was 86 to 182 keV. In a hot-rod and a uniform phantom imaging study, this energy window achieved better image quality than the other two energy window settings: open-energy window, and 120 to 160 keV. In the open-energy window setting, the lower energy threshold is set to a minimal value above detector electronic noise, and the upper threshold is open.
AB - We developed add-on SPECT imaging capability on an animal PET scanner to implement a cost effective dual modality imaging system. Using PET detectors, designed for 511 keV photons, to acquire relatively low energy gamma rays from SPECT radionuclides has some inherent downsides, including the sub-optimal energy resolution and background radioactivity from 176Lu, a long-lived radioisotope of Lutetium that is present in most of the modern PET detector scintillators. To address these downsides, this study was carried out to optimize the detector energy window setting in SPECT imaging mode. A LabPET-8 scanner with a pinhole collimator insert was studied through a Monte Carlo simulation package. The energy spectra of detector crystal's intrinsic 176Lu, and a cylindrical imaging phantom's 99mTc emissions were simulated and assessed. The optimal energy window setting was derived by searching the lower and upper energy thresholds that maximize the noise-equivalent-count rate (NECR), an index adapted from its PET counterpart. The characteristics of 176Lu energy spectrum obtained from simulation matches that reported in the literature and reproduces the features of 176Lu energy spectra measured on the LabPET-8 scanner. The optimal energy window setting derived from the simulation was 86 to 182 keV. In a hot-rod and a uniform phantom imaging study, this energy window achieved better image quality than the other two energy window settings: open-energy window, and 120 to 160 keV. In the open-energy window setting, the lower energy threshold is set to a minimal value above detector electronic noise, and the upper threshold is open.
UR - https://www.scopus.com/pages/publications/84904192888
U2 - 10.1109/NSSMIC.2013.6829104
DO - 10.1109/NSSMIC.2013.6829104
M3 - Conference contribution
SN - 9781479905348
T3 - IEEE Nuclear Science Symposium Conference Record
BT - 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2013
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 2013 60th IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2013
Y2 - 27 October 2013 through 2 November 2013
ER -