Skip to main navigation Skip to search Skip to main content

Fracture of the neck of a lateral flare femoral implant in total hip arthroplasty

  • Adam P. Breceda
  • , Michael J. Derogatis
  • , Patrick Lee
  • , Vincent J. Vigorita
  • , Eli Bryk
  • , Paul S. Issack

Research output: Contribution to journalArticlepeer-review

Abstract

We present the case of a 38-year-old man who presented 7 years after primary total hip replacement with a fracture of the neck of a lateral flare femoral stem and catastrophic polyethylene wear. The unique design of the lateral flare hip stem has been shown previously to be associated with accelerated polyethylene wear, whereas the stem remains well fixed. The resultant polyethylene wear results in the harder cobalt chrome head abrading the titanium shell generating metallic debris. This may have resulted in proximal migration of the implant neck into the cup with subsequent neck-cup impingement and implant fracture.

Original languageEnglish
Pages (from-to)73-77
Number of pages5
JournalJournal of Long-Term Effects of Medical Implants
Volume28
Issue number2
DOIs
StatePublished - 2018

Keywords

  • Lateral flare
  • Metallosis
  • Stem fracture
  • Trunnion fracture

Fingerprint

Dive into the research topics of 'Fracture of the neck of a lateral flare femoral implant in total hip arthroplasty'. Together they form a unique fingerprint.

Cite this