Skip to main navigation Skip to search Skip to main content

Guidewire Dislodgment of Inferior Vena Cava Filters During Insertion of Central Venous Catheters

  • Steven S. Kang
  • , Marc A. Borge
  • , M. Ashraf Mansour
  • , Nicos Labropoulos
  • , William H. Baker
  • Loyola University Medical Center

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

The authors report two cases of titanium Greenfield filters dislodged by guidewires during the insertion of central venous catheters. Both were trauma victims who had the venous catheters inserted into femoral veins several days after correct placement of the inferior vena cava filters. A literature review found nine others cases of guidewire dislodg ment of vena cava filters. Eight Vena-Tech filters and one titanium Greenfield filter had been in place for up to 2 months. The guidewires were inserted into femoral, subclavian, internal jugular, or axillary veins. Filters were dislodged cephalad into the atrium, brachiocephalic, and subclavian veins, or caudad into the distal inferior vena cava. Limiting the distance that a guidewire is inserted into a vein during central line insertion is the easiest way to avoid this complication. Inserting the straight end of the guidewire would also be effective.

Original languageEnglish
Pages (from-to)587-593
Number of pages7
JournalVascular and Endovascular Surgery
Volume31
Issue number5
DOIs
StatePublished - Sep 1997

Fingerprint

Dive into the research topics of 'Guidewire Dislodgment of Inferior Vena Cava Filters During Insertion of Central Venous Catheters'. Together they form a unique fingerprint.