Abstract
A reversed-phase isocratic high-performance liquid chromatographic method has been developed for the determination of AG-331, a novel thymidylate synthase inhibitor, in human serum and urine. The method involves a solid-phase extraction from C18 cartridges without addition of an internal standard. The methanol eluent is evaporated under nitrogen at 40°C, and reconstituted in mobile phase, acetonitrile-water (35:65, v/v) containing 25 mM ammonium phosphate. Separation of AG-331 was obtained on a C18 column at a flow-rate of 1 ml/min. Chromatographic signals were monitored by a photodiode-array detector at a primary wavelength of 457 nm with a bandwidth of 4.8 nm. Standard curves are linear in the range of 22-2175 ng/ml in plasma and 44-2175 ng/ml in urine, respectively. The extraction recovery ranged from 92.9-102.4%. Intra-day coefficient of variation was less than 9.5%, and inter-day coefficient of variation was less than 14.3% for an AG-331 concentration of 44 ng/ml. This method has been used to characterize the pharmacokinetics of AG-331 in cancer patients as part of ongoing Phase I trials.
| Original language | English |
|---|---|
| Pages (from-to) | 307-314 |
| Number of pages | 8 |
| Journal | Journal of Chromatography B: Biomedical Sciences and Applications |
| Volume | 666 |
| Issue number | 2 |
| DOIs | |
| State | Published - Apr 21 1995 |
Fingerprint
Dive into the research topics of 'High-performance liquid chromatographic method for the determination of AG-331, a novel anti-cancer agent, in human serum and urine using solid-phase extraction and photodiode-array detection'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver