Measuring the presampled MTF from a reduced number of flat-field images using the noise response (NR) method

  • Canon Inc.
  • SUNY Buffalo

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

4 Scopus citations

Abstract

We evaluate a new method for measuring the presampled modulation transfer function (MTF) using the noise power spectrum (NPS) obtained from a few flat-field images acquired at one exposure level. The NPS is the sum of structure, quantum, and additive instrumentation noise, which are proportional to exposure squared, exposure, and a constant, respectively, with the spatial-frequency dependence of the quantum noise depending partly on the detector MTF. Cascaded linear-systems theory was used to derive an exact and generic relationship that was used to isolate noise terms and enable determination of the MTF directly from the noise response, thereby circumventing the need for precision test objects (slit, edge, etc.) as required by standard techniques. Isolation of the quantum NPS by fitting the total NPS versus exposure obtained using 30 flat-field images each at six or more different exposure levels with a linear regression provides highly accurate MTFs. A subset of these images from indirect digital detectors was used to investigate the accuracy of measuring the MTF from 30 or fewer flat-field images obtained at a single exposure level. Analyzing as few as two images acquired at a single exposure resulted in no observable systematic error. Increasing the number of images analyzed resulted in an increase in accuracy. Fifteen images provided comparable accuracy with the most rigorous slope approach, with less than 5% variability, suggesting additional image acquisitions may be unnecessary. Reducing the number of images acquired for the noise response method further simplifies and facilitates routine MTF measurements.

Original languageEnglish
Title of host publicationMedical Imaging 2011
Subtitle of host publicationPhysics of Medical Imaging
DOIs
StatePublished - 2011
EventMedical Imaging 2011: Physics of Medical Imaging - Lake Buena Vista, FL, United States
Duration: Feb 13 2011Feb 17 2011

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume7961

Conference

ConferenceMedical Imaging 2011: Physics of Medical Imaging
Country/TerritoryUnited States
CityLake Buena Vista, FL
Period02/13/1102/17/11

Keywords

  • MTF
  • NPS
  • SSXII
  • detector
  • flat panel detector
  • image quality
  • performance
  • quality assurance
  • two-dimensional MTF

Fingerprint

Dive into the research topics of 'Measuring the presampled MTF from a reduced number of flat-field images using the noise response (NR) method'. Together they form a unique fingerprint.

Cite this