Skip to main navigation Skip to search Skip to main content

Skin dose mapping for non-uniform x-ray fields using a backscatter point spread function

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

6 Scopus citations

Abstract

Beam shaping devices like ROI attenuators and compensation filters modulate the intensity distribution of the xray beam incident on the patient. This results in a spatial variation of skin dose due to the variation of primary radiation and also a variation in backscattered radiation from the patient. To determine the backscatter component, backscatter point spread functions (PSF) are generated using EGS Monte-Carlo software. For this study, PSF's were determined by simulating a 1 mm beam incident on the lateral surface of an anthropomorphic head phantom and a 20 cm thick PMMA block phantom. The backscatter PSF's for the head phantom and PMMA phantom are curve fit with a Lorentzian function after being normalized to the primary dose intensity (PSFn). PSFn is convolved with the primary dose distribution to generate the scatter dose distribution, which is added to the primary to obtain the total dose distribution. The backscatter convolution technique is incorporated in the dose tracking system (DTS), which tracks skin dose during fluoroscopic procedures and provides a color map of the dose distribution on a 3D patient graphic model. A convolution technique is developed for the backscatter dose determination for the nonuniformly spaced graphic-model surface vertices. A Gafchromic film validation was performed for shaped x-ray beams generated with an ROI attenuator and with two compensation filters inserted into the field. The total dose distribution calculated by the backscatter convolution technique closely agreed with that measured with the film.

Original languageEnglish
Title of host publicationMedical Imaging 2017
Subtitle of host publicationPhysics of Medical Imaging
EditorsTaly Gilat Schmidt, Joseph Y. Lo, Thomas G. Flohr
PublisherSPIE
ISBN (Electronic)9781510607095
DOIs
StatePublished - 2017
EventMedical Imaging 2017: Physics of Medical Imaging - Orlando, United States
Duration: Feb 13 2017Feb 16 2017

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume10132

Conference

ConferenceMedical Imaging 2017: Physics of Medical Imaging
Country/TerritoryUnited States
CityOrlando
Period02/13/1702/16/17

Keywords

  • Backscatter
  • Compensation filter
  • Dose
  • EGSnrc
  • Point spread function
  • ROI attenuator

Fingerprint

Dive into the research topics of 'Skin dose mapping for non-uniform x-ray fields using a backscatter point spread function'. Together they form a unique fingerprint.

Cite this