Skip to main navigation Skip to search Skip to main content

Ultrashort echo time MRI of cortical bone at 7 Tesla field strength: A feasibility study

  • Roland Krug
  • , Peder Eric Zufall Larson
  • , Chunsheng Wang
  • , Andrew J. Burghardt
  • , Douglas A.C. Kelley
  • , Thomas M. Link
  • , Xiaoliang Zhang
  • , Daniel B. Vigneron
  • , Sharmila Majumdar
  • University of California at San Francisco
  • GE Healthcare United States

Research output: Contribution to journalArticlepeer-review

36 Scopus citations

Abstract

Purpose: To implement and examine the feasibility of a three-dimensional (3D) ultrashort TE (UTE) sequence on a 7 Tesla (T) clinical MR scanner in comparison with 3T MRI at high isotropic resolution. Materials and Methods: Using an in-house built saddle coil at both field strengths we have imaged mid-diaphysial sections of five fresh cadaveric specimens of the distal tibia. An additional in vivo scan was performed at 7 Tesla using a quadrature knee coil. Results: Using the same type of saddle coil at both field strengths, a significant increase in SNR at 7T compared with 3T (factor 1.7) was found. Significantly shorter T2* values were found at the higher field strength (T2* = 552.2 ± 126 μs at 7T versus T2* = 1163 ± 391 μs at 3T). Conclusion: UHF MRI at 7T has great potential for imaging tissues with short T2.

Original languageEnglish
Pages (from-to)691-695
Number of pages5
JournalJournal of Magnetic Resonance Imaging
Volume34
Issue number3
DOIs
StatePublished - Sep 2011

Keywords

  • 7 Tesla
  • cortical bone
  • ultra-short MRI

Fingerprint

Dive into the research topics of 'Ultrashort echo time MRI of cortical bone at 7 Tesla field strength: A feasibility study'. Together they form a unique fingerprint.

Cite this