Skip to main navigation Skip to search Skip to main content

Where to B in dZ/dt

  • Ohio State University

Research output: Contribution to journalArticlepeer-review

221 Scopus citations

Abstract

The B point on the impedance cardiograph waveform corresponds to the opening of the aortic valve and is an important parameter for calculating systolic time intervals, stroke volume, and cardiac output. Identifying the location of the B point is sometimes problematic because the characteristic upstroke that serves as a marker of this point is not always apparent. Here is presented a reliable method for B point identification, based on the consistent relationship between the R to B interval (RB) and the interval between the R-wave and the peak of the dZ/dt function (RZ). The polynomial function relating RB to RZ (RB=1.233RZ-0.0032RZ2-31.59) accounts for 90%-95% of the variance in the B point location across ages and gender and across baseline and stress conditions. This relation affords a rapid approximation to B point measurement that, in noisy or degraded signals, is superior to visual B point identification and to a derivative-based estimate.

Original languageEnglish
Pages (from-to)113-119
Number of pages7
JournalPsychophysiology
Volume44
Issue number1
DOIs
StatePublished - Jan 2007

Keywords

  • B point
  • ECG
  • Impedance cardiography
  • LVET
  • PEP
  • Reliability
  • Validity